3-MCPD & GE Rapid Quantification Kit



Background - The Problem

Glycidol and its esters (GE) and 3-chloropropane-1,2-diol (3-MCPD) are foodborne contaminants that can be formed during the processing of various fatty acid rich foodstuffs. Due to the inherent risk posed by 3-MCPD and GE, regulatory bodies and current industry roadmaps aim to reduce the levels of 3-MCPD below 2 ppm and GE below 1ppm by September 2020.



Dose dependent Health Risks (3-MCPD and GE)

  • Progressive Nephrotoxicity
  • Lowers male fertility
  • Carcinogenic Potential

Amongst foodstuffs, edible oils, specifically, refined palm oil contains the highest levels of 3-MCPD and GE.

Class Foods 3-MCPD (mg/kg) GE (mg/kg)
Oils
Refined palm oil 1.1-10.0 0.30-10
Refined coconut oil 1.4-1.7 n.a
Refined hazelnut/walnut oil 1.2-19.0 0.5-1.4
Refined olive oil <0.3-2.5 n.a


Our Solution: The Camtech Rapid Quantification Kit




Rapid Test Kit for Screening 3-Chloropropane-1,2-Diol (3-MCPD) and Glycidol Ester (GE) in Refined Cooking Oil

This rapid assay kit includes a colorimetric assay that quantifies levels of 3-MCPD and Total (3-MCPD + GE) in Edible oils. A major source of 3-MCPD fatty acid esters in consumer food products. Independent levels of GE are then also determined via a simple calculations. The rapid test kit requires less than 2 hours per test batch (4 samples) from sample preparation to result at an approximate cost of ~$50 per sample. In comparison, standard GC-MS analysis services can cost around $300 per sample with a turnaround time of 2-7 days.


Summary

  • Quantification of 3-MCPD and GE in Edible oils
  • Can detect free 3-MCPD + GE and their esters
  • Simple sample preparation
  • Fast turnaround (<13 min/sample)
  • LOQ (0.1 ppm)
  • Ideal for batch screening and process monitoring
  • Cost-effective alternative ($50/sample) vs GCMS ($300)
  • Colorimetric response in visible range
  • Online calculator for instant results
  • Expert technical support available for new customers

Accuracy


Calibration curves for the 3-MCPD and GE assays done with Camtech’s kit and compared to corn oil spiked with 3-MCPD-1,2-bis-palmitoyl ester and glycidyl oleate (n=3; Error Bars = SD). There is a high correlation between the kit and the spiked samples for 3-MCPD (R2=0.994) and GE (R2=0.994).



Comparison of 3-MCPD and GE calibration curves from measurements done with Camtech’s Kit or GC-MS using spiked samples. A high correlation for both 3-MCPD (R2=0.99), and GE (R2=0.98) between the kit and GC-MS results has been observed. (Kit: n=3, error bars=SD; GC-MS: n=1, error bars=SD)



Corn oil was used as base oil and spiked with low (0.5mg/kg), medium (2.0mg/kg), high (4.0 mg/kg) concentrations of 3-MCPD-1,2-bis-palmitoyl ester and glycidyl oleate. The current LOQ’s for both 3-MPCD and GE fall within the range of the EU regulations for edible oils. (n=8 for each graph; error bars=SD) GE was determined indirectly using the 3-MCPD and Total Values.



Edible oil samples from various factories (n=4) were analysed using in-house GC-MS systems and compared to in-house Camtech kit values. Across the concentration range the kit shows a high degree of correlation for 3-MCPD (R2=0.98) and Total (R2=0.99) (n=16)


Proficiency Test



Results from the Fapas®Food Chemistry Proficiency Test #2658 for 3-MCPD esters & Glycidyl Esters. The performance of Camtech (Lab #64) is indicated by the red line. The results indicate Camtech had minimum deviation from the average result for 3-MCPD (z-score: -0.2) & GE (z-score: 0.6) (n=101).


Kit components:

Each kit is sufficient for 50 assays.


Chemicals provided

  • Solvent: (100 ml)
  • Reagent 1 (300 ml)
  • Reagent 2: (130 ml)
  • Reagent A: (25 ml)
  • Reagent B: (25 ml) Store at 4°C
  • Reagent C: (50 ml) Store at 4°C
  • Disposables provided on request
  • 15/50 ml centrifuge tubes (100 pcs.)


Equipment required but not provided

  • 2.5, 100 to 1000 μl pipettes with tips
  • Vortex
  • Oven/ water bath at 37 °C
  • Boiling water bath/kettle
  • Centrifuge (15 ml tubes | 4000 rpm)
  • Spectrophotometer/ colorimeter (min 0.01 resolution)


Precautions and disclaimer

This product is for lab use only, not for drug production, household, or other uses. Please read the MSDS before using the kit.



Storage conditions and shelf life

Items marked 4°C must be stored in a refrigerator. Place other components at room temperature away from direct sunlight. Kit is valid for 3 months from date of shipping.



Tested & Compatible Oils:

  • Palm oil and its fractions
  • Stearin
  • Sunflower Oil
  • Soybean oil
  • Rice bran oil
  • Canola oil
  • Mixed vegetable fats
  • Margarine fat
  • Baking and confectionery fat